会员体验
专利管家(专利管理)
工作空间(专利管理)
风险监控(情报监控)
数据分析(专利分析)
侵权分析(诉讼无效)
联系我们
交流群
官方交流:
QQ群: 891211   
微信请扫码    >>>
现在联系顾问~
热词
    • 3. 发明授权
    • Dual-mode pulse oximeter
    • 双模脉搏血氧仪
    • US07530949B2
    • 2009-05-12
    • US10911391
    • 2004-08-03
    • Ammar Al AliDon CarothersDavid DalkeMohamed K. DiabJulian GoldmanMassi E. KianiMichael LeeJerome NovakRobert SmithVal E. Vaden
    • Ammar Al AliDon CarothersDavid DalkeMohamed K. DiabJulian GoldmanMassi E. KianiMichael LeeJerome NovakRobert SmithVal E. Vaden
    • A61B5/00
    • A61B5/14551A61B5/0002A61B5/1495A61B5/742A61B2560/0271A61B2560/0456A61B2562/085G06F19/00
    • A pulse oximeter has an integrated mode in which it operates as a plug-in module for a multiparameter patient monitoring system (MPMS). The pulse oximeter also has a portable mode in which operates separately from the MPMS as a battery-powered handheld or standalone instrument. The pulse oximeter has a sensor port that receives a photo-plethysmographic signal as input to an internal processor. The pulse oximeter processes this sensor signal to derive oxygen saturation and pulse rate measurements. In the portable mode, this information is provided on its display, and stored in memory for trend capability. In the integrated mode, the pulse oximeter provides oxygen saturation and pulse rate measurements to the MPMS through a docking station to be displayed on a MPMS monitor. In the integrated mode, the portable pulse oximeter docks to the docking station, which in turn is inserted in one or more MPMS slots. The docking station can function as a simple electrical pass-through device between the docked portable pulse oximeter and the MPMS or it can provide a MPMS communications interface.
    • 脉搏血氧计具有集成模式,其作为用于多参数病人监测系统(MPMS)的插件模块。 脉搏血氧计还具有便携式模式,其中与MPMS分开操作,作为电池供电的手持或​​独立仪器。 脉搏血氧计具有传感器端口,其接收作为对内部处理器的输入的光学体积描记信号。 脉冲血氧计处理该传感器信号以导出氧饱和度和脉搏率测量。 在便携式模式中,该信息在其显示器上提供,并存储在存储器中用于趋势能力。 在集成模式中,脉搏血氧计通过对接站向MPMS提供氧饱和度和脉搏率测量,以在MPMS监视器上显示。 在集成模式中,便携式脉搏血氧仪停靠在对接站,对接站又插入一个或多个MPMS槽中。 对接站可以作为对接便携式脉搏血氧计与MPMS之间的简单电气穿越装置,或者可以提供MPMS通信接口。
    • 4. 发明授权
    • Pulse oximeter probe-off detector
    • 脉搏式血氧计探测器
    • US07471969B2
    • 2008-12-30
    • US10721607
    • 2003-11-25
    • Mohamed K. DiabAmmar Al Ali
    • Mohamed K. DiabAmmar Al Ali
    • A61B5/1455
    • A61B5/7221A61B5/14551A61B5/6843A61B2560/0276
    • A processor provides signal quality based limits to a signal strength operating region of a pulse oximeter. These limits are superimposed on the typical gain dependent signal strength limits. If a sensor signal appears physiologically generated, the pulse oximeter is allowed to operate with minimal signal strength, maximizing low perfusion performance. If a sensor signal is potentially due to a signal induced by a dislodged sensor, signal strength requirements are raised. Thus, signal quality limitations enhance probe off detection without significantly impacting low perfusion performance. One signal quality measure used is pulse rate density, which defines the percentage of time physiologically acceptable pulses are occurring. If the detected signal contains a significant percentage of unacceptable pulses, the minimum required signal strength is raised proportionately. Another signal quality measure used in conjunction with pulse rate density is energy ratio, computed as the percentage of total energy contained in the pulse rate fundamental and associated harmonics.
    • 处理器为脉冲血氧计的信号强度操作区域提供基于信号质量的限制。 这些极限叠加在典型的增益相关信号强度极限上。 如果传感器信号出现生理生成,脉搏血氧计允许以最小的信号强度进行操作,最大限度地降低灌注性能。 如果传感器信号潜在地由于由移动的传感器引起的信号,则会提高信号强度要求。 因此,信号质量限制增强了探针检测,而不会显着影响低灌注性能。 使用的一种信号质量测量是脉率密度,其定义了生理上可接受的脉冲发生的时间的百分比。 如果检测到的信号含有显着百分比的不可接受的脉冲,则所需的最小信号强度成比例地提高。 与脉搏密度结合使用的另一个信号质量测量是能量比,计算为脉冲速率基波和相关谐波中包含的总能量的百分比。