会员体验
专利管家(专利管理)
工作空间(专利管理)
风险监控(情报监控)
数据分析(专利分析)
侵权分析(诉讼无效)
联系我们
交流群
官方交流:
QQ群: 891211   
微信请扫码    >>>
现在联系顾问~
热词
    • 2. 发明授权
    • Implantable cardioverter defibrillator having a smaller mass
    • 植入式心律转复除颤器质量较小
    • US5957956A
    • 1999-09-28
    • US963154
    • 1997-11-03
    • Mark W. KrollTheodore P. AdamsKenneth M. AndersonCharles U. Smith
    • Mark W. KrollTheodore P. AdamsKenneth M. AndersonCharles U. Smith
    • A61N1/375A61N1/39H01M2/10H01M6/50H01M16/00A61N1/18
    • A61N1/375A61N1/3956A61N1/3758H01M16/00H01M2/1066H01M6/5016
    • A capacitor-discharge implantable cardioverter defibrillator (ICD) has a relatively smaller mass of less than about 120 grams. The smaller mass of the ICD is achieved by selecting and arranging the internal components of the ICD to deliver a maximum defibrillation countershock optimized in terms of a minimum physiologically effective current (I.sub.pe), rather than a minimum defibrillation threshold energy (DFT). As a result of the optimization in terms of a minimum effective current I.sub.pe, there is a significant decrease in the maximum electrical charge energy (E.sub.c) that must be stored by the capacitor of the ICD to less than about 30 Joules, even though a higher safety margin is provided for by the device. Due to this decrease in the maximum E.sub.c, as well as corollary decreases in the effective capacitance value required for the capacitor and the net energy storage required of the battery, the overall displacement volume of the ICD is reduced to the point where subcutaneous implantation of the device in the pectoral region of human patients is practical. The size of the capacitor is reduced because the effective capacitance required can be less than about 125 .mu.F. By optimizing both the charging time and the countershock duration for the smaller maximum E.sub.c, the size of the battery is reduced because the total energy storage capacity can be less than about 1.0 Amp-hours. In the preferred embodiment, the charging time for each defibrillation countershock is reduced to less than about 10 seconds and the pulse duration of the countershock is reduced to less than about 6 milliseconds.
    • 电容放电植入式心律转复除颤器(ICD)的质量相对较小,小于约120克。 通过选择和布置ICD的内部部件来实现ICD的较小质量,以提供根据最小生理有效电流(Ipe)而不是最小除颤阈值能量(DFT)优化的最大除颤抗倒伏。 作为最小有效电流Ipe优化的结果,ICD的电容器必须存储的最大电荷能量(Ec)显着降低到小于约30焦耳,即使较高的 设备提供安全余量。 由于最大Ec的这种减小以及电容器所需的有效电容值和电池所需的净能量存储的必然减小,ICD的总位移体积减少到皮下植入 在人类胸腔的器官是实用的。 电容器的尺寸减小,因为所需的有效电容可以小于约125μF。通过优化较小的最大Ec的充电时间和反冲时间,电池的尺寸减小,因为总能量存储容量 可以小于约1.0安培小时。 在优选实施例中,每个去纤维颤动反射的充电时间减少到小于约10秒,并且该反冲的脉冲持续时间减小到小于约6毫秒。
    • 3. 发明授权
    • Implantable cardioverter defibrillator having a smaller energy storage
capacity
    • 具有较小能量储存能力的植入式心律转复除颤器
    • US5827326A
    • 1998-10-27
    • US412920
    • 1995-03-29
    • Mark W. KrollTheodore P. AdamsKenneth M. AndersonCharles U. Smith
    • Mark W. KrollTheodore P. AdamsKenneth M. AndersonCharles U. Smith
    • A61N1/375A61N1/378A61N1/39H01M2/10H01M6/50H01M16/00
    • A61N1/375A61N1/3787A61N1/3931A61N1/3956A61N1/3975H01M16/00A61N1/3758A61N1/3906A61N1/3912H01M2/1066H01M6/5016
    • A capacitor-discharge implantable cardioverter defibrillator (ICD) has a relatively smaller energy storage capacity of less than about 1.0 Amp-hours. The smaller energy storage capacity of the ICD is achieved by selecting and arranging the internal components of the ICD to deliver a maximum defibrillation countershock optimized in terms of a minimum physiologically effective current (I.sub.pe) rather than a minimum defibrillation threshold energy (DFT). As a result of the optimization in terms of a minimum effective current I.sub.pe, there is a significant decrease in the maximum electrical charge energy (E.sub.c) that must be stored by the capacitor of the ICD to less than about 30 Joules, even though a higher safety margin is provided for by the device. Due to this decrease in the maximum E.sub.c, as well as corollary decreases in the effective capacitance value required for the capacitor and the net energy storage required of the battery, the overall displacement volume of the ICD is reduced to the point where subcutaneous implantation of the device in the pectoral region of human patients is practical. The size of the capacitor is reduced because the effective capacitance required can be less than about 125 .mu.F. By optimizing both the charging time and the countershock duration for the smaller maximum E.sub.c, the size of the battery is reduced because the total energy storage capacity can be less than about 1.0 Amp-hours. In the preferred embodiment, the charging time for each defibrillation countershock is reduced to less than about 10 seconds and the pulse duration of the countershock is reduced to less than about 6 milliseconds.
    • 电容放电植入式心律转复除颤器(ICD)具有小于约1.0安培小时的相对较小的储能能力。 通过选择和布置ICD的内部部件来实现ICD的较小的能量存储容量,以提供根据最小生理有效电流(Ipe)而不是最小除颤阈值能量(DFT)优化的最大除颤反相。 作为最小有效电流Ipe优化的结果,ICD的电容器必须存储的最大电荷能量(Ec)显着降低到小于约30焦耳,即使较高的 设备提供安全余量。 由于最大Ec的这种减小以及电容器所需的有效电容值和电池所需的净能量存储的必然减小,ICD的总位移体积减少到皮下植入 在人类胸腔的器官是实用的。 电容器的尺寸减小,因为所需的有效电容可以小于约125μF。通过优化较小的最大Ec的充电时间和反冲时间,电池的尺寸减小,因为总能量存储容量 可以小于约1.0安培小时。 在优选实施例中,每个去纤维颤动反射的充电时间减少到小于约10秒,并且该反冲的脉冲持续时间减小到小于约6毫秒。
    • 5. 发明授权
    • Apparatus for generating biphasic waveforms in an implantable
defibrillator
    • 用于在可植入除颤器中产生双相波形的装置
    • US5871505A
    • 1999-02-16
    • US457307
    • 1995-06-01
    • Theodore P. AdamsMark W. Kroll
    • Theodore P. AdamsMark W. Kroll
    • A61N1/39
    • A61N1/3956A61N1/3906A61N1/3912
    • A method and apparatus for generating biphasic waveforms for an implantable defibrillator uses two disparate capacitor systems. A first phase of the biphasic waveform is generated from the charge stored in a first capacitor system, and a second phase of less energy than the first phase is generated from the charge stored in a second capacitor system. By using a second capacitor system for the second phase of the biphasic waveform that stores less energy than the first capacitor system, the overall size and power requirements of the implantable device can be reduced. In a preferred embodiment, the disparate capacitor systems have different effective capacitances, with the second capacitor system preferably having a higher effective capacitance, but a lower charging voltage, than the first capacitor system.
    • 用于产生可植入除颤器的双相波形的方法和装置使用两个不同的电容器系统。 从存储在第一电容器系统中的电荷产生双相波形的第一相,并且从存储在第二电容器系统中的电荷产生比第一相少的能量的第二相。 通过使用存储比第一电容器系统更少的能量的双相波形的第二相的第二电容器系统,可以减小可植入装置的总体尺寸和功率要求。 在优选实施例中,不同的电容器系统具有不同的有效电容,其中第二电容器系统优选地具有比第一电容器系统更高的有效电容,但是具有较低的充电电压。
    • 6. 发明授权
    • Implantable cardioverter defibrillator having a smaller displacement
volume
    • 具有较小排量的植入式心律转复除颤器
    • US5697953A
    • 1997-12-16
    • US582529
    • 1996-01-11
    • Mark W. KrollTheodore P. AdamsKenneth M. AndersonCharles U. Smith
    • Mark W. KrollTheodore P. AdamsKenneth M. AndersonCharles U. Smith
    • A61N1/375A61N1/39H01M2/10H01M6/50H01M16/00
    • A61N1/375A61N1/3956A61N1/3758H01M16/00H01M2/1066H01M6/5016
    • A capacitor-discharge implantable cardioverter defibrillator (ICD) has a relatively longer device life of greater that 5 years. The longer life of the ICD is achieved by selecting and arranging the internal components of the ICD to deliver a maximum defibrillation countershock optimized in terms of a minimum physiologically effective current (I.sub.pe), rather than a minimum defibrillation threshold energy (DFT). As a result of the optimization in terms of a minimum effective current I.sub.pe, there is a significant decrease in the maximum electrical charge energy (E.sub.c) that must be stored by the capacitor of the ICD to less than about 30 Joules, even though a higher safety margin is provided for by the device. Due to this decrease in the maximum E.sub.c, as well as corollary decreases in the effective capacitance value required for the capacitor and the net energy storage required of the battery, the overall displacement volume of the ICD is reduced to the point where subcutaneous implantation of the device in the pectoral region of human patients is practical. The size of the capacitor is reduced because the effective capacitance required can be less than about 125 .mu.F. By optimizing both the charging time and the countershock duration for the smaller maximum E.sub.c, the size of the battery is reduced because the total energy storage capacity can be less than about 1.0 Amp-hours. In the preferred embodiment, the charging time for each defibrillation countershock is reduced to less than about 10 seconds and the pulse duration of the countershock is reduced to less than about 6 milliseconds.
    • 电容放电可植入式心律转复除颤器(ICD)具有相对较长的器件寿命,大于5年。 通过选择和布置ICD的内部组件来实现ICD的更长寿命,以提供根据最小生理有效电流(Ipe)而不是最小除颤阈值能量(DFT)优化的最大除颤抗倒伏。 作为最小有效电流Ipe优化的结果,ICD的电容器必须存储的最大电荷能量(Ec)显着降低到小于约30焦耳,即使较高的 设备提供安全余量。 由于最大Ec的这种减小以及电容器所需的有效电容值和电池所需的净能量存储的必然减小,ICD的总位移体积减少到皮下植入 在人类胸腔的器官是实用的。 电容器的尺寸减小,因为所需的有效电容可以小于约125μF。通过优化较小的最大Ec的充电时间和反冲时间,电池的尺寸减小,因为总能量存储容量 可以小于约1.0安培小时。 在优选实施例中,每个去纤维颤动反射的充电时间减少到小于约10秒,并且该反冲的脉冲持续时间减小到小于约6毫秒。
    • 10. 发明授权
    • Cardiac arrhythmia detection using interdependent multiple parameter
settings
    • 使用相互依赖的多参数设置进行心律失常检测
    • US5458620A
    • 1995-10-17
    • US195557
    • 1994-02-14
    • Theodore P. AdamsMark W. Kroll
    • Theodore P. AdamsMark W. Kroll
    • A61N1/362A61N1/39
    • A61N1/3621A61N1/36585A61N1/3956Y10S128/925
    • A programmable implantable cardioverter defibrillator (ICD) provides signals from a plurality of cardiac sensing modalities to a control circuit within the ICD that analyzes the signals and determines whether to deliver an electrical countershock therapy. An interdependent detection parameter threshold is programmably established for each of a programmably selectable combination of two or more cardiac sensing modalities by selecting at least two corner points that will define a boundary condition for a given cardiac dysrhythmia. Whenever the threshold is exceeded, the control circuit automatically diagnoses a cardiac dysrhythmia and delivers a preprogrammed electrical countershock therapy regimen for the diagnosed dysrhythmia. Multiple interdependent detection parameter thresholds may be combined in a variety of ways. More than one combination of two or more cardiac sensing modalities may be used to define a multivariant boundary condition for the given cardiac dysrhythmia. Alternatively, more than one set of corner points for a particular combination of cardiac sensing modalities may be used so as to create multiple detection envelopes, one for each different type of cardiac dysrhythmia. Finally, both of these options may be programmed together so as to allow for the definition of multivariant performance envelopes for different cardiac dysrhythmias.
    • 可编程植入式心律转复除颤器(ICD)将来自多个心脏感测模式的信号提供给ICD内的控制电路,该控制电路分析信号并确定是否递送电的临时治疗。 通过选择将限定给定的心脏心律失常的边界条件的至少两个角点,可编程地针对两个或更多个心脏感测模态的可编程选择的组合来为每个可编程地建立相互依赖的检测参数阈值。 每当超过阈值时,控制电路自动诊断心律失常并提供用于诊断的心律失常的预编程电褥疮治疗方案。 可以以多种方式组合多个相互依赖的检测参数阈值。 可以使用两种或多种心脏感测模式的多于一种组合来定义给定的心脏心律失常的多变量边界条件。 或者,可以使用用于心脏感测模态的特定组合的多于一组角点,以便产生多个检测包络,每个不同类型的心脏心律失常一个。 最后,这两个选项可以一起编程,以便允许定义用于不同心脏心律失常的多变量性能包络。